Cleartext Transmission of Sensitive Information
CVE-2019-6540
Summary
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.
- LOW
- ADJACENT_NETWORK
- NONE
- UNCHANGED
- NONE
- NONE
- HIGH
- NONE
CWE-319 - Cleartext Transmission of Sensitive Information
The software transmits sensitive or security-critical data in cleartext in a communication channel that can be sniffed by unauthorized actors.
References
Advisory Timeline
- Published