Skip to main content

Missing Authentication for Critical Function

CVE-2016-8355

Severity High
Score 9.9/10

Summary

An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. CADD-Solis Medication Safety Software grants an authenticated user elevated privileges on the SQL database, which would allow an authenticated user to modify drug libraries, add and delete users, and change user permissions. According to Smiths-Medical, physical access to the pump is required to install drug library updates.

  • LOW
  • NETWORK
  • HIGH
  • CHANGED
  • NONE
  • LOW
  • HIGH
  • HIGH

CWE-306 - Missing Authentication for Critical Function

The software does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of resources.

References

Advisory Timeline

  • Published