Skip to main content

Use of Hard-coded Credentials

CVE-2015-3953

Severity High
Score 9.8/10

Summary

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

  • LOW
  • NETWORK
  • HIGH
  • UNCHANGED
  • NONE
  • NONE
  • HIGH
  • HIGH

CWE-798 - Use of Hard-coded Credentials

The software contains hard-coded credentials, such as a password or cryptographic key, which it uses for its own inbound authentication, outbound communication to external components, or encryption of internal data.